A reliable, compact instrument designed for student laboratory use. Supplied with operation manual, fuse, power cord and 2 probes.
CRT |
|
Type: |
6” rectangular, internal graticule |
Display Area: |
8 ´ 10 DIV (DIV=10mm) |
Accelerating Voltage: |
1.9kV |
Brightness and Focusing: |
Controlled at the front panel |
Trace Rotation: |
Adjusted at the front panel |
Vertical System |
|
Sensitivity: |
5mV – 20V/DIV, ±; 1mV/DIV – 1V/DIV, ±5% (5 times, only CH1) 12 calibration steps |
Width of Band |
DC (AC: 10Hz) –20MHz |
Rise Time |
<17.5ns |
Input Impedance |
1MOhm ± 3%; 30pF ±5pF |
Input Coupling |
AC, GND, DC |
Max. Input Voltage: |
400V (DC + AC peak) |
Vertical Mode: |
CH1, CH2, DUAL (CHOP, ALT), ADD, CH2 inverse |
CH1 Signal Output: |
25mV/DIV 50 Ohm; 20HZ-10MHz (-3dB) |
Horizontal System |
|
Time Factor: |
0.2ms – 0.2s/DIV 19 steps in order 1-2-5 Multiplication factor 10 times (Max. sweep speed 20ns/DIV) |
Accuracy: |
±3%, ±5% (10 times) |
Trigger System |
|
Mode: |
AUTO, NORM, TV-V, TV-H |
Source: |
VERT(DUAL, ALT), CH1, EXT, LINE |
Coupling: |
AC |
Polarity: |
Positive or Negative |
External Triggering |
|
Signal Input: |
Input impedance: 1Mohm ±3%+ 25pF±5pF Max. input voltage: 400V |
Sensitivity Internal |
20Hz – 2MHz: 0.5 DIV (0.2V p-p) 2MHz – 20MHz: 1.5 DIV (0.8 p-p) TV: 1.0 DIV (1V p-p) 20MHz – 40MHz: 1.5 DIV (0.8 p-p) |
X-Y Operation |
|
Input: |
Axis X: CH1 input signal Axis Y: CH2 input signal |
Width of Band |
Axis X: DC – 500kHz |
Phase Error: |
<3° (DC – 50kHz) |
Axis Z |
|
Input Impedance: |
20kOhm – 30kOhm |
Input Signal: |
5V p-p (Negative signal Highlighted) |
Width of Band: |
DC – 2MHz |
Max. Input Voltage: |
30V (DC + AC peak) |
Calibration Source: |
1kHz 0.5V p-p Square wave |
Power Source: |
110-127 VAC, 220-240VAC, 50/60Hz |
Dimensions: |
410 ´ 316 ´ 132mm |
Weight: |
7.8kg |
Tender Specifications are :